FDA-Device2023-10-04Class II
Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.
Hazard
Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set
Sold states
US Nationwide distribution in the states of NJ, WI.
Affected count
1344 pumps
Manufactured in
50 High St Ste 50, N/A, North Andover, MA, United States
Products
Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2635-2023More Fresenius Kabi USA, LLC recalls
- FDA-Drug2026-06-17Fresenius Kabi epinephrine injection recalled for failed impurity specifications
- FDA-Drug2026-04-15Fresenius Kabi sodium chloride injection recalled for sterility assurance
- FDA-Drug2026-04-15Fresenius Kabi sodium chloride injection recalled for sterility assurance
- FDA-Drug2026-04-15Fresenius Kabi sodium chloride injection recalled for sterility concerns
- FDA-Drug2026-04-15Fresenius Kabi sodium chloride injection recalled for lack of sterility assurance
Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief