FDA-Device2017-07-05Class II
Exactech 1.5" Novation Calcar Planer Guide Tip To assist the surgeon in the implantation of Novation hip system components according to a conventional technique for total hip replacement. Specifically, these instruments are reusable, surgically invasive, and intended for transient use. ech 1.5" Novation Calcar Planer Guide Tip
Exactech, Inc.
Hazard
The firm determined that the labeling (surgical technique) should be updated to clarify the proper technique and instrumentation to perform calcar planing to the bone, i.e., to include a caution statement about the potential for instrument breakage if bending or torsional forces are applied.
Sold states
Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and to the states of : AL, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, LA, MA, MD, ME, MN, NC, NJ, NV, NY, OH, OK, OR, RI, SC, TN, TX, VA, WA, WI, Hawaii., and to the countries of : Argentina, Australia, Austria, Brazil, China, Colombia, France, Germany, Guatemala, India, Italy, Japan, Spain, Switzerland, Turkey, and United Kingdom
Affected count
477
Manufactured in
2320 NW 66th Ct, N/A, Gainesville, FL, United States
Products
Exactech 1.5" Novation Calcar Planer Guide Tip To assist the surgeon in the implantation of Novation hip system components according to a conventional technique for total hip replacement. Specifically, these instruments are reusable, surgically invasive, and intended for transient use. ech 1.5" Novation Calcar Planer Guide Tip
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2644-2017Don't want to check this manually?