FDA-Device2014-09-24Class II

AXIOM Luminos dRF system with SW VD10 or on Luminos dRF Max The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.

Siemens Medical Solutions USA, Inc
Hazard

It was determined that a potential malfunction may occur on the Luminos dRF system with SW VD10 or on Luminos dRF Max when the table side control is exposed to fluids. Fluid may infiltrate the tableside control panel and result in malfunction. In the very unlikely case of fluid infiltration a permanent short circuit within the control panel of the Luminos dRF / Luminos dRF Max will automatically d

Sold states
Nationwide Distribution including IL VA TN PA TX IN MN NM GA MI FL CA NY ME OK MO NJ and AZ.
Affected count
33
Manufactured in
51 Valley Stream Pkwy, N/A, Malvern, PA, United States
Products
AXIOM Luminos dRF system with SW VD10 or on Luminos dRF Max The Axiom Luminos dRF is intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic studies. Using a digital flat detector, it can perform a range of applications including general R/F, angiography and pediatric examinations. The Axiom Luminos dRF is a device intended to visualize anatomical structures by converting a pattern of X-ray into a visible image.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2649-2014