FDA-Device2023-10-11Class II

Venue Go Standard Carts Ref: (H45181VC and H45103VCW), used with Venue Go R2, R3, R4 systems R2, R3, and R4 systems; pulsed doppler imaging system

GE MEDICAL SYSTEMS, ISRAEL LTD.
Hazard

Some Venue Go Standard Carts can develop an internal failure of the vertical/horizontal tilt adjustment mechanism which can result in the Venue Go system detaching from the cart and falling.

Sold states
Worldwide distribution - US Nationwide distribution including government and military distribution as well as PR.
Affected count
5986 devices
Manufactured in
Nativ Haa??or Street no. 1, N/A, HAIFA, N/A, Israel
Products
Venue Go Standard Carts Ref: (H45181VC and H45103VCW), used with Venue Go R2, R3, R4 systems R2, R3, and R4 systems; pulsed doppler imaging system

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2652-2023

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief