FDA-Device2020-07-29Class II
Nuvasive NVM5 EMG Module, REF 8020015, Rx ONLY, CE2797, Nonsterile, UDI: 00887517206633 - Product Usage: is intended for intraoperative neurophysiologic monitoring during spinal surgery.
NuVasive Inc
Hazard
The NVM5 EMG Module kit intended for intraoperative neurophysiologic monitoring during spinal surgery contains incorrect component, part number 1715945, SSEP harness instead of the the correct part number 1747065, NVM5 Harness, Gen 3.5 EM. This issue can result in delay in therapy or lack of EMG neuromonitoring during surgery if the surgeon elects to proceed without the NVM5.
Sold states
US Nationwide distribution including in the states of AR, CA, IA, IL, WI. OUS: None
Affected count
23
Manufactured in
7475 Lusk Blvd, N/A, San Diego, CA, United States
Products
Nuvasive NVM5 EMG Module, REF 8020015, Rx ONLY, CE2797, Nonsterile, UDI: 00887517206633 - Product Usage: is intended for intraoperative neurophysiologic monitoring during spinal surgery.
Official notice