FDA-Device2024-08-28Class II
DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
Beckman Coulter Mishima K.K.
Hazard
Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.
Sold states
Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL, MI, MN, MS, NE, NY, OR, PA and SC. The countries of Lebanon, South Africa.
Affected count
24 analyzers
Manufactured in
454-32, Higashino; Nagaizumi-Cho, Sunto-Gun, Japan
Products
DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2691-2024Don't want to check this manually?