FDA-Device2016-09-07Class II
BD Vacutainer Push Button Blood Collection Set, Sterile, Rx only, 0.6 x 19 mm x 305 mm, 23G x 3/4" x 12", REF 367342.
Becton Dickinson & Company
Hazard
Becton Dickinson is recalling various lot numbers of BD Vacutainer Push Button Blood Collection Set because of reports that specific lots exhibit separation of fron and rear barrels upon activation of the safety feature, which retracts the needle.
Sold states
US and International
Affected count
1,788,150 units
Manufactured in
1 Becton Dr, N/A, Franklin Lakes, NJ, United States
Products
BD Vacutainer Push Button Blood Collection Set, Sterile, Rx only, 0.6 x 19 mm x 305 mm, 23G x 3/4" x 12", REF 367342.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2710-2016Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief