FDA-Device2020-08-12Class I
Alaris System PC Unit Model 8000, a modular infusion pump and monitoring system
CareFusion 303, Inc.
Hazard
Cleaning agents may damage Inter-Unit Interface (IUI) Connectors, which may lead to interruption of communication or power between PC Unit and modules, which may result in an infusion that stops with a CHANNEL DISCONNECTED message, and an audio/visual alarm on the PC Unit and an interruption of therapy or monitoring.
Sold states
Worldwide distribution. US nationwide including Guam, Puerto Rico, American Samoa, and Military Pacific. Countries of: TR, QA, CA, AE, KW, MX, GI, IL, SA, CH, NL, BE, TW, ZA, PH, CN, AU, IN, FR, MY, NZ, GB, KR, PK, DE, SG, JP, CU, IT, and BH
Affected count
37141
Manufactured in
10020 Pacific Mesa Blvd, San Diego, CA, United States
Products
Alaris System PC Unit Model 8000, a modular infusion pump and monitoring system
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2716-2020Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief