FDA-Device2016-09-14Class II

ArjoHuntleigh Sara Combilizer; Intended to facilitate early immobilization, rehabilitation, and care of patients.

ARJOHUNTLEIGH POLSKA Sp. z.o.o.
Hazard

ArjoHuntleigh has concluded that a device tipping hazard may occur on the current Sara Combilizer lifting mechanism, providing a potential for risk to the patient or the caregiver. The tipping phenomenon has been found during internal device testing. To date, there have been no adverse events or customer complaints related to the tipping phenomenon. The tipping hazard may occur when fully loaded Sara Combilizer in "stretcher mode" is lifted to the highest position (approximately 984 mm = 38 3/4") and tilted backwards to the angle of 25 degrees¿-28 degrees¿. Although such scenario has been defined as highly unlikely to occur during use with patients, as a commitment to quality, ArjoHuntleigh is providing customers with information on action needed. The Sara Combilizer may stay in use until the upgrade will be performed, on the condition that the tilt angle in the device ''stretcher mode" will be no greater than minus 15 degrees¿ and the device is not loaded over 330 lbs (150 kg).

Sold states
CO, AL, MA, NC, MI, NY, FL, MD, CA, VA, CT, TX, and IL; Australia, Austria, Denmark, Germany, Israel, Japan, Norway, Sweden, Switzerland, UAE, and UK.
Affected count
106
Manufactured in
UI. KS. PIOTRA WAWRZYNIAKA 2, KOMORNIKI, Poland
Products
ArjoHuntleigh Sara Combilizer; Intended to facilitate early immobilization, rehabilitation, and care of patients.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2718-2016

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief