FDA-Device2018-09-26Class II
Temperature sensor Foley catheter 8FR, Catalogue Number 102205100880AB Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
Degania Silicone, Ltd.
Hazard
The catheters are suspected to contain defective sensors; if used, defective catheter will show lower body temperature than the actual body temperature of a patient.
Sold states
Worldwide Distribution - US Nationwide in the states: IL, NC, NY and the countries of : Belgium, Israel, Japan, and Mexico.
Affected count
11488
Manufactured in
Degania Bet, N/A, Dganya Bet, N/A, Israel
Products
Temperature sensor Foley catheter 8FR, Catalogue Number 102205100880AB Product Usage: This is sterile, single use, two-way silicone Foley Catheter with a thermistor embedded in the third lumen. It is intended for use in the drainage/ collection of urine from the urinary bladder and simultaneous monitoring of the body core temperature during surgery or post-surgical period.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2755-2018Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief