FDA-Device2016-09-14Class II

Coronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Coronary Artery Perfusion Cannula with Balloon The Cannula consists of a single lumen vinyl tube provided for with a vinyl female luer connector bonded proximally and a vinyl cuff bonded distally. The cuff is pre-molded to its inflated configuration and is purposely positioned directly over a hole generated in the extruded tube for purposes of receiving the infused cardioplegia solution. These devices come in direct conta

Sorin Group USA, Inc.
Hazard

The tip of the cannula should be bent at either a 45 or 90 degree angle. Samples have been returned where the angle of the tip is less than the requirement.

Sold states
Worldwide Distribution - US including IL, MO, TX. Internationally to Canada and Italy.
Affected count
1205 units worldwide
Manufactured in
14401 W 65th Way, Arvada, CO, United States
Products
Coronary Artery Perfusion Cannula with Balloon (part number CP-2200X). The Coronary Artery Perfusion Cannula with Balloon The Cannula consists of a single lumen vinyl tube provided for with a vinyl female luer connector bonded proximally and a vinyl cuff bonded distally. The cuff is pre-molded to its inflated configuration and is purposely positioned directly over a hole generated in the extruded tube for purposes of receiving the infused cardioplegia solution. These devices come in direct contact with the central circulatory system but they are not intended to control, diagnose, monitor or correct a defect.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2757-2016