FDA-Device2017-08-02Class II
Pacific Hemostasis FDP (Fibrinogen Degradation Products) Assay Kit (30 Determinations), REF/Model 100650, UDI 00845275000863, IVD; --- Fisher Diagnostics, a division of Fisher Scientific Company LLC ,a part of Thermo Fisher Scientific Inc., Middletown, VA 22645-1905 USA --- For determination of Fibrinogen Degradation products in serum and urine.
Fisher Diagnostics
Hazard
An issue was discovered with the Pacific Hemostasis Fibrinogen Degradation Products (FDP) kit, list number 100650, lot 948546 and FDP Latex Reagent, list 100651, lot 890199. While the kit performs accurately at FDP concentrations greater than 20 ¿g/ml, there is an observed reduction in sensitivity when FDP levels are less than 20 ¿g/ml. This reduction in sensitivity has been tied to FDP Latex Reagent lot 890199 and may result in false negatives.
Sold states
USA (nationwide) including Puerto Rico, and Internationally to Colombia
Affected count
331 units
Manufactured in
8365 Valley Pike, N/A, Middletown, VA, United States
Products
Pacific Hemostasis FDP (Fibrinogen Degradation Products) Assay Kit (30 Determinations), REF/Model 100650, UDI 00845275000863, IVD; --- Fisher Diagnostics, a division of Fisher Scientific Company LLC ,a part of Thermo Fisher Scientific Inc., Middletown, VA 22645-1905 USA --- For determination of Fibrinogen Degradation products in serum and urine.