FDA-Device2016-09-14Class II

Innocor Cardiopulmonary Exercise and Lung Function Testing System Non-invasive diagnostic device that measures or calculates a number of hemodynamic and metabolic parameters.

Innovision A/S
Hazard

The purpose of the recall is to replace data transmission cable (USB-cable) connecting the lnnocor device with an external computer. The replacement USB cable will provide galvanic isolation of the lnnocor from the external computer via opto-couplers, as opposed to the grounding protection currently in use. The replacement is intended to prevent risk of electric shock.

Sold states
Worldwide Distribution -- US, to Massachusetts and Washington, DC; and, the countries of Canada, Germany, and United Kingdom.
Affected count
8
Manufactured in
Skovvaenget 2, Glamsbjerg, Denmark
Products
Innocor Cardiopulmonary Exercise and Lung Function Testing System Non-invasive diagnostic device that measures or calculates a number of hemodynamic and metabolic parameters.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2769-2016