FDA-Device2016-09-21Class III
LDX Optical Check Cassettes (OCC) Model Number: 10-228 and 14-535 Cassette in box with foam insert. The cassette is used to check the optics system of the LDX analyzer. Product Usage: The Optics Check Cassette is used to check the optics system of the Alere Cholestech LDX¿ Analyzer. It is run once a day before patient samples are tested and after the analyzer has been moved or serviced.
Alere San Diego, Inc.
Hazard
Incorrect expiration date on the magnetic strip that does not match the labeled expiration date.
Sold states
Worldwide Distribution - U.S Nationwide in the states of: TN, MA, GA, OK, VA, FL, TN, MO, AL, TX, IL, NY, VT, RI, KY, WI, AL, NC, LA, SD, TX, VA, MI, MN, IN, IA, CA, AR, PA, HI, WV, KS, CO, CT, UT and the countries of Japan, England, Canada
Affected count
12,951
Manufactured in
9975 Summers Ridge Rd, N/A, San Diego, CA, United States
Products
LDX Optical Check Cassettes (OCC) Model Number: 10-228 and 14-535 Cassette in box with foam insert. The cassette is used to check the optics system of the LDX analyzer. Product Usage: The Optics Check Cassette is used to check the optics system of the Alere Cholestech LDX¿ Analyzer. It is run once a day before patient samples are tested and after the analyzer has been moved or serviced.