FDA-Device2020-08-19Class II

Bioseal Trachea Extender 1/pl 50pk/Cs, REF TRAX01/50, LOT XXXX, STERILE EO, Single Use Only, Date of Manufacture XXXX, Expiration Date XXXX, LF, UDI: (01) 00630094916019 - Product Usage: Airway connectors intended to connect a breathing gas source to a tracheal tube, tracheostomy tube, or mask.

Bioseal Corporation
Hazard

The reported stress crack in the port of the swivel elbow where the flip cap is inserted could result in leakage from the breathing system resulting in the prescribed ventilation not being delivered to the patient. These cracks could possibly leak if they reach a significant size and pose a serious heath risk to patient.

Sold states
US Nationwide distribution including in the state of CA.
Affected count
900 units
Manufactured in
167 W Orangethorpe Ave, N/A, Placentia, CA, United States
Products
Bioseal Trachea Extender 1/pl 50pk/Cs, REF TRAX01/50, LOT XXXX, STERILE EO, Single Use Only, Date of Manufacture XXXX, Expiration Date XXXX, LF, UDI: (01) 00630094916019 - Product Usage: Airway connectors intended to connect a breathing gas source to a tracheal tube, tracheostomy tube, or mask.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2795-2020