FDA-Device2020-08-19Class II

uMI 550 PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Shanghai United Imaging Healthcare Co., Ltd.
Hazard

1) Service function problem may cause false marking of a bad channel resulting in ring artifacts; 2) Potential intermittent scout scanning interruption due to angle signal drift caused by external interference; 3) Metal edge overlap of mylar strip in gantry may create sharp edge. May lead to repeat scanning and abrasions (N/A for Serial No. 200036)

Sold states
US Nationwide distribution including in the state of Texas.
Affected count
4 units
Manufactured in
No. 2258 Chengbei Rd, Jiading Ind., N/A, Shanghai, N/A, China
Products
uMI 550 PET/CT Diagnostic Imaging System - Product Usage: This system is intended to be operated by qualified healthcare professionals to assist in the detection, localization, diagnosis, staging, restaging, treatment planning and treatment response evaluation for diseases and disorders in, but not limit to, oncology, cardiology and neurology.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2800-2020