FDA-Device2020-08-26Class II

NobelReplace Conical Connection (CC) PMC RP 4.3x10mm, REF: 37291 - Product Usage: implants are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function.

NOBEL BIOCARE SERVICES AG
Hazard

Dental implant cap label may incorrectly indicate implant measurements 5.0x8mm. Actual implant measurements are 4.3x10mm. If not noticed prior to implant placement, this could lead to: 1) A second surgical procedure; 2) Implant replacement; 3) If placed in posterior lower jaw, the implant may damage the inferior alveolar nerve canal or perforate a small artery requiring immediate implant removal.

Sold states
Worldwide distribution - U.S. Nationwide distribution including in the states of FL and NY and the countries of Albania and Hong Kong.
Affected count
188
Manufactured in
Balz-zimmermannstrasse 7, N/A, Kloten, N/A, Switzerland
Products
NobelReplace Conical Connection (CC) PMC RP 4.3x10mm, REF: 37291 - Product Usage: implants are intended to be used in the upper or lower jaw bone (osseointegration) and used for anchoring or supporting tooth replacements to restore chewing function.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2828-2020