FDA-Device2018-08-29Class II
Connecting Bolt, Model 14-442093
Zimmer Biomet, Inc.
Hazard
Product was manufactured from the incorrect material, which shows a potential decrease in the fatigue strength of the connecting bolt that can lead to fracture.
Sold states
AR, AZ, CA, FL, GA, IN, KY, LA, MA, MI, MN, MO, MT, NJ, NY, OH, SC, TX, UT & WI
Affected count
247 units
Manufactured in
56 E Bell Dr, Warsaw, IN, United States
Products
Connecting Bolt, Model 14-442093
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2875-2018Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief