FDA-Device2018-09-05Class II
Cortisol Saliva Luminescence Immunoassay The IBL International Cortisol Saliva Luminescence immunoassay is intended for the in-vitro diagnostic quantitative determination of Cortisol in human saliva and for use as an aid in the diagnosis and treatment of adrenal disorders. The device is not intended for point-of-care settings.
Tecan US, Inc.
Hazard
Product is not meeting specification for traceability to NIST cortisol standard, generating 1.3-1.8 times higher cortisol values with laboratory samples in comparison to NIST cortisol standard.
Sold states
Worldwide distribution; US distribution to states of: GA, IA, LA, MA, MN, OR, PA, SC, and UT; and countries of: Germany, Finland, Switzerland, France, Austria, Denmark, Poland, Norway, Belgium, Netherlands, Italy, Canada, South Africa, Russia, and Japan.
Affected count
870 (USA) 180 (OUS)
Manufactured in
9401 Globe Center Dr Ste 140, N/A, Morrisville, NC, United States
Products
Cortisol Saliva Luminescence Immunoassay The IBL International Cortisol Saliva Luminescence immunoassay is intended for the in-vitro diagnostic quantitative determination of Cortisol in human saliva and for use as an aid in the diagnosis and treatment of adrenal disorders. The device is not intended for point-of-care settings.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2923-2018Don't want to check this manually?