FDA-Device2018-09-05Class II

N Latex CDT Kit

CSL Behring GmbH
Hazard

Siemens Healthcare Diagnostics has observed a negative bias for the carbohydrate-deficient transferrin (CDT) measurement when using the affected N Latex CDT lots in comparison to the HPLC method. CDT results in absolute concentrations may be influenced by patients transferrin levels and, therefore, results are reported as ratio of CDT to total transferrin, called %CDT. As the transferrin determination with N Antiserum to Human Transferrin is not affected, the observed effect leads to calculated %CDT values that show a negative bias of approximately 15% compared to the HPLC method. This could result in a shift of weak positive patient samples into the reference range of 1.19  2.47 %CDT which was derived from a study population of healthy adults. There are no control or calibrator materials that would catch this bias.

Sold states
Domestic - DE, IN, KS, MO & VA Internationally - Australia Austria Belgium Canada China Czech Republic Finland France Germany Hungary India Italy Japan Latvia Lithuania Netherlands Norway Poland Portugal Saudi Arabia Slovakia Slovenia South Africa South Koria Spain Sweden Switzerland Taiwan Turkey United Kingdom Uruguay
Affected count
6635 units
Manufactured in
Emil-von-Behring-Str. 76, N/A, Marburg, N/A, Germany
Products
N Latex CDT Kit

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2931-2018

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief