FDA-Device2018-09-05Class II

Haemonetics OrthoPAT System-Integrated Set. Product Code: 1150H-00 OrthoPAT orthopedic perioperative autotransfusion system indicated for use to salvage RBCs from blood lost intraoperatively and postoperatively during surgical procedures

Haemonetics Corporation
Hazard

OrthoPAT¿ reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the collected blood cannot be processed.

Sold states
US Nationwide Distribution
Affected count
432 units
Manufactured in
400 Wood Rd, N/A, Braintree, MA, United States
Products
Haemonetics OrthoPAT System-Integrated Set. Product Code: 1150H-00 OrthoPAT orthopedic perioperative autotransfusion system indicated for use to salvage RBCs from blood lost intraoperatively and postoperatively during surgical procedures

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2950-2018

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief