FDA-Device2018-09-05Class II
Haemonetics OrthoPAT System-Integrated Set. Product Code: 1150H-00 OrthoPAT orthopedic perioperative autotransfusion system indicated for use to salvage RBCs from blood lost intraoperatively and postoperatively during surgical procedures
Haemonetics Corporation
Hazard
OrthoPAT¿ reservoirs are missing the cannula inside the reservoirs. The absence of this cannula prevents the emptying of blood from the reservoirs. The inability to empty the reservoir means that the collected blood cannot be processed.
Sold states
US Nationwide Distribution
Affected count
432 units
Manufactured in
400 Wood Rd, N/A, Braintree, MA, United States
Products
Haemonetics OrthoPAT System-Integrated Set. Product Code: 1150H-00 OrthoPAT orthopedic perioperative autotransfusion system indicated for use to salvage RBCs from blood lost intraoperatively and postoperatively during surgical procedures
Official notice