FDA-Device2024-09-11Class II
Sevoflurane E_Vap.(Safety Filling adapter) used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number: 115-066756-00
Mindray DS USA, Inc. dba Mindray North America
Hazard
V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
Sold states
US Distribution to states of: California, Georgia, Kentucky, Mississippi, New Jersey, Puerto Rico, Texas, and West Virginia; and International (Foreign) to: Canada
Affected count
96 units
Manufactured in
800 Macarthur Blvd, Mahwah, NJ, United States
Products
Sevoflurane E_Vap.(Safety Filling adapter) used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentration to an A9 anesthesia system. Part Number: 115-066756-00
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2950-2024Don't want to check this manually?