FDA-Device2024-09-18Class I

Pneupac paraPAC plus 300 ventilator kit, REF P300NXX

Smiths Medical ASD Inc.
Hazard

There is a potential for inadvertent tidal volume knob movement from the original setting when set at high (1000-1500 mL) and low (70-150 mL) settings.

Sold states
Worldwide
Affected count
2198 units
Manufactured in
6000 Nathan Ln N, Minneapolis, MN, United States
Products
Pneupac paraPAC plus 300 ventilator kit, REF P300NXX

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-2973-2024