FDA-Device2020-10-07Class II

ImmuTest Drugs of Abuse Cup (AMT Item Number/ AMT Description): IMCA-5AB-W / "ImmuTest Drugs of Abuse Cup AMP1000+BZO300+COC300+OPI2000+THC50 With Cr, Ni, pH, Bl, S.G.", IMC-8BBCE/ "ImmuTest Drugs of Abuse Cup 8 Drug Detection: (COC300/THC50)+AMP1000+BZO300+BUP10+MET500+MTD300+OPI300 NO ADULTERATION", IMCA-8CM5 /"ImmuTest Drugs of Abuse Cup (OXY100/MDMA500)+(PCP25/AMP1000)+COC150+MET500+OPI2000+THC50 with Cr, Ni, pH, Bl, S.G."

Ameditech Inc
Hazard

Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.

Sold states
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Affected count
27,375 units
Manufactured in
9940 Mesa Rim Rd, N/A, San Diego, CA, United States
Products
ImmuTest Drugs of Abuse Cup (AMT Item Number/ AMT Description): IMCA-5AB-W / "ImmuTest Drugs of Abuse Cup AMP1000+BZO300+COC300+OPI2000+THC50 With Cr, Ni, pH, Bl, S.G.", IMC-8BBCE/ "ImmuTest Drugs of Abuse Cup 8 Drug Detection: (COC300/THC50)+AMP1000+BZO300+BUP10+MET500+MTD300+OPI300 NO ADULTERATION", IMCA-8CM5 /"ImmuTest Drugs of Abuse Cup (OXY100/MDMA500)+(PCP25/AMP1000)+COC150+MET500+OPI2000+THC50 with Cr, Ni, pH, Bl, S.G."

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3036-2020