FDA-Device2017-08-23Class II

HomeChoice Pro Automated PD system Model numbers: 5C8310, 5C8310R Product Usage: The HomeChoice Automated Personal Cycler is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Baxter Healthcare Corporation
Hazard

Baxter Healthcare Corporation has been made aware that users may not be following the instructions in the Operators Manual and incorrectly (using sharp objects) opening disposable set packaging while setting up their Peritoneal Dialysis (PD) therapy, damaging the cassettes for the HomeChoice or HomeChoice PRO cyclers. If damaged cassettes are used, the cyclers may not consistently detect very small holes/cuts in the sheeting of the cassette in the patient valve region and the cycler may deliver air into the patient.

Sold states
Worldwide Distribution
Affected count
23732 units
Manufactured in
1 Baxter Pkwy, N/A, Deerfield, IL, United States
Products
HomeChoice Pro Automated PD system Model numbers: 5C8310, 5C8310R Product Usage: The HomeChoice Automated Personal Cycler is intended for automatic control of dialysate solution exchanges in the treatment of pediatric and adult renal failure patients undergoing peritoneal dialysis.

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3041-2017