FDA-Device2020-10-07Class II
ToxCup Drug Screen Cup (AMT Item Number/ AMT Description): HT29A/"ToxCup¿ Drug Screen Cup COC300/MET500/OPI300/THC50/AMP1000/PCP25/OXY100 Adulteration Screen for CR - NI - PH - BL - S.G.", DT14/ "ToxCup¿ Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100; PCP25/PPX300", HT15/ "ToxCup¿ Drug Screen Cup AMP1000/COC300/MET500/PCP25/ THC50/OPI2000", HT15A/ ToxCup¿ Drug Screen Cup AMP1000/COC300/MET500/PCP25/ THC50/OPI2000 Ad
Ameditech Inc
Hazard
Various immunoassay enzymes failed product performance, thereby the products do not consistently detect all calibrator and secondary compounds at the levels stated in the labeling.
Sold states
US - Nationwide OUS - Australia, France, Germany, New Zealand, Singapore and Spain
Affected count
189,175 units
Manufactured in
9940 Mesa Rim Rd, N/A, San Diego, CA, United States
Products
ToxCup Drug Screen Cup (AMT Item Number/ AMT Description): HT29A/"ToxCup¿ Drug Screen Cup COC300/MET500/OPI300/THC50/AMP1000/PCP25/OXY100 Adulteration Screen for CR - NI - PH - BL - S.G.", DT14/ "ToxCup¿ Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100; PCP25/PPX300", HT15/ "ToxCup¿ Drug Screen Cup AMP1000/COC300/MET500/PCP25/ THC50/OPI2000", HT15A/ ToxCup¿ Drug Screen Cup AMP1000/COC300/MET500/PCP25/ THC50/OPI2000 Adulteration Screen for CR - NI - PH - BL - S.G.", PT15/ "ToxCup¿ Drug Screen Cup AMP1000/COC300/MET500/OPI300/THC50" DT13A/ "ToxCup¿ Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100 w/ Cr, Ni, pH, BI, S.G.", DT14A "ToxCup¿ Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; BUPG10/TCA1000; MDMA500/OXY100; PCP25/PPX300 w/ Cr, Ni, pH, Bl, S.G. Finished Devices", DT12A/"ToxCup¿ Drug Screen Cup THC50/COC150; OPI300/MET500; AMP500/BZO300; BAR300/MTD300; MDMA500/OXY100; PCP25/PPX300 w/ Cr, Ni, pH, BI, S.G.", PT12A/ "ToxCup¿ Drug Screen Cup COC300/MET500/OPI2000/PCP25/THC50 Adulteration Screen for CR - NI - PH - BL - S.G.", PT11A/ "ToxCup¿ Drug Screen Cup AMP1000/COC300/OPI2000/PCP25/THC50 Adulteration Screen for CR - NI - PH - BL - S.G.", HT30/ "ToxCup¿ Drug Screen Cup AMP1000/COC300/MET1000/PCP25/THC50/ OPI2000"
Official notice