FDA-Device2017-09-06Class II
mcompass 2-channel Balloon Catheters The mcompass Biofeedback Anorectal Manometry System is for use on patients requiring anorectal pressure studies. mcompass Anorectal Manometry System (RMD-001-001) which includes a FOB component connected to the balloon catheter to inflate and deflate and a medical grade tablet PC to run the software. The software is not included in the recall, only the balloon catheter component.
Medspira, Llc
Hazard
Potential failure in the balloon bond in the inner catheter stem, leading to the separation of the balloon from the inner stem, leaving it in the rectal cavity.
Sold states
Worldwide Distribution - US Distribution to the states of :CA, FL, IA, ID, and TX., and to the countries of : Austria, Italy, Malaysia, Philippines, Sweden, and United Kingdom
Affected count
355 catheters (255 US - 100 foreign. )
Manufactured in
2718 Summer St NE, N/A, Minneapolis, MN, United States
Products
mcompass 2-channel Balloon Catheters The mcompass Biofeedback Anorectal Manometry System is for use on patients requiring anorectal pressure studies. mcompass Anorectal Manometry System (RMD-001-001) which includes a FOB component connected to the balloon catheter to inflate and deflate and a medical grade tablet PC to run the software. The software is not included in the recall, only the balloon catheter component.
Official notice