FDA-Device2017-09-20Class I
Skintact Electrode for Defibrillation, Pacing, Monitoring, Cardioversion. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only. Used with automated external defibrillators
Leonhard Lang Medizintechnik GmbH
Hazard
Certain lots of the defibrillation electrodes model DF59N and DF59NC may have a delay in connecting or not connect at all with the defibrillator ZOLL AED PLUS. This may cause a situation in which a patient, who is in a life threatening condition and requires a defibrillation shock, cannot be treated in good time.
Sold states
Worldwide distribution. US distribution in the states of : FL & TX. Austria, China, France, Germany, Greece, Hungary, India, Italy, Panama, Philippines, Poland, Saudi Arabia, Serbia, South Africa, South Korea, Spain, United Arab Emirates, United Kingdom.
Affected count
8,249 units total
Manufactured in
Archenweg 56, N/A, Innsbruck, N/A, Austria
Products
Skintact Electrode for Defibrillation, Pacing, Monitoring, Cardioversion. Product Usage: Multifunction electrode for external defibrillation, pacing, cardioversion, and monitoring. The device is non-sterile and for single use only. Used with automated external defibrillators
Official notice