FDA-Device2017-10-04Class II
Zimmer Periarticular Locking Plate System Locking Screw Tap Standard Quick-Connect, 2.7 mm Diameter, Item Number/EDI 00236015327, Nonsterile. orthopedic surgical instrument.
Zimmer Biomet, Inc.
Hazard
Various trauma and extremity instruments (drill bits and screw taps) were manufactured from raw material with an anomaly. The anomaly may be on or below the surface and can increase the risk of instrument fracture.
Sold states
Nationwide in USA; Internationally to: Australia, Brazil, Canada, China, El Salvador, Germany, Guatemala, India, Japan, Panama, Malaysia, Mexico, Netherlands, South Korea, Singapore, Taiwan & Trinidad.
Affected count
171 units
Manufactured in
1800 W Center St, N/A, Warsaw, IN, United States
Products
Zimmer Periarticular Locking Plate System Locking Screw Tap Standard Quick-Connect, 2.7 mm Diameter, Item Number/EDI 00236015327, Nonsterile. orthopedic surgical instrument.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3149-2017Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief