FDA-Device2024-09-25Class II

Marodyne-LiV, part number 3045-133, Low intensity vibration exerciser supporting osteoporosis treatment

Btt Health Gmbh
Hazard

Some power supply units in Austria and Germany had a voltage drop in the millisecond range, which led to a cancellation of the treatment and subsequent restart of the medical device. The treatment is only restarted by a step-off/step-on (step-on for activation).

Sold states
US Nationwide distribution in the states of Florida.
Affected count
62 units
Manufactured in
Billerberg 7, Inning, Germany
Products
Marodyne-LiV, part number 3045-133, Low intensity vibration exerciser supporting osteoporosis treatment

Official notice

https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3162-2024

Don't want to check this manually?

We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.

Get the Sunday Brief