FDA-Device2018-09-26Class II
VITEK(R) 2 AST-P655, For susceptibility testing of Staphylococcus spp. Enterococcus spp., ad S agalactiae against specified antimicrobials, REF 421913 Product Usage: VITEK¿ 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK¿ 2 utilizes growth-based biochemical patterns to determine identification. The VITEK¿ 2 provides minimal inhibitory concentration (MIC) r
bioMerieux, Inc.
Hazard
Routine internal review of Device History Records identified six VITEK(R) 2 AST card lots were manufactured with less drug (gentamicin) than required for the Gentamicin High-Level Resistance (ghlr01n) screen test due to an error when making 1500mL and 2000mL media well preparations (solutions).
Sold states
Worldwide Distribution US Nationwide and the countries of: Turkey, Austria, Switzerland, Czech Republic, Germany, Netherlands,
Affected count
10919 cartons
Manufactured in
100 Rodolphe St, N/A, Durham, NC, United States
Products
VITEK(R) 2 AST-P655, For susceptibility testing of Staphylococcus spp. Enterococcus spp., ad S agalactiae against specified antimicrobials, REF 421913 Product Usage: VITEK¿ 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. The VITEK¿ 2 utilizes growth-based biochemical patterns to determine identification. The VITEK¿ 2 provides minimal inhibitory concentration (MIC) results for most organism/drug combinations as well as a category interpretation (S, I, or R) consistent with accepted criteria.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3180-2018Don't want to check this manually?