FDA-Device2018-09-26Class II
PATHFAST NTproBNP Catalog Number: PF1061-KUS PATHFAST¿ NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST¿ for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole blood and plasma.
Polymedco, Inc.,/Polymedco Cancer Diagnostic Products, Llc
Hazard
A pinhole caused by the laser printer in the aluminum sheet over the ALP reagent wells caused leakage through the pinhole. If the cartridge with leakage is used, higher measurement value as low as +12% and no higher than +20% bias may be obtained.
Sold states
US Distribution to states of: AR, CA, FL, ID, LA, MN, NY and TX.
Affected count
272 KITS in total
Manufactured in
510 Furnace Dock Rd, N/A, Cortlandt Manor, NY, United States
Products
PATHFAST NTproBNP Catalog Number: PF1061-KUS PATHFAST¿ NTproBNP is a product for in-vitro diagnostic use with the in vitro diagnostic system PATHFAST¿ for the quantitative measurement of N-terminal-pro B-type Natriuretic Peptide (NT-proBNP) in heparinized or EDTA whole blood and plasma.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3186-2018Don't want to check this manually?