FDA-Device2017-10-04Class II
UniPerc(R) Adjustable Flange Tracheostomy Tube kit, Reorder Number 100/897/080 CZ The UniPerc(R) Tracheostomy Tubes can be used during the surgical placement of a stoma or as a replacement tube to an already existing stoma for patients with larger necks. The tracheostomy tubes are removed and may be replaced after no more than 30 days with an identical device. The range may include the following configurations: cuffed or uncuffed tracheostomy tubes, tracheostomy tube inner cannula, cleaning
Smiths Medical ASD Inc.
Hazard
Smiths Medical became aware that one lot of the 8.0mm UniPerc(R) Adjustable Flange Tracheostomy Tube kit, product reorder number 100/897/080 CZ, lot number 3308971, contains an incorrect 9.0mm sized obturator instead of the correct 8.0mm sized obturator.
Sold states
INTERNATIONAL ONLY: GB, FR, AT, DE, IT, SE, IE, BE, ES, LU
Affected count
89 units
Manufactured in
6000 Nathan Ln N, N/A, Minneapolis, MN, United States
Products
UniPerc(R) Adjustable Flange Tracheostomy Tube kit, Reorder Number 100/897/080 CZ The UniPerc(R) Tracheostomy Tubes can be used during the surgical placement of a stoma or as a replacement tube to an already existing stoma for patients with larger necks. The tracheostomy tubes are removed and may be replaced after no more than 30 days with an identical device. The range may include the following configurations: cuffed or uncuffed tracheostomy tubes, tracheostomy tube inner cannula, cleaning brushes and tracheostomy tube holder.
Official notice
https://www.accessdata.fda.gov/scripts/ires/index.cfm?Product=Z-3213-2017Don't want to check this manually?
We email you a Sunday digest of new federal recalls relevant to parents — free, no credit card. Or upgrade to as-it-happens alerts for the brands you actually own.
Get the Sunday Brief