FDA-Drug2012-06-27Class II
LifeGas OXYGEN COMPRESSED UN1072 USP, Distributed by: Linde
Linde Gas LLC
Hazard
CGMP Deviations: The pressure gages, vacuum gages, and thermometer had surpassed the calibration expiry period, which may have resulted in overfill/underfill of oxygen cylinders.
Sold states
Distributed only within Ohio. No Federal Government, foreign or Canadian consignees received product.
Affected count
283 cylinders
Manufactured in
7029 Huntley Rd Ste K, N/A, Columbus, OH, United States
Products
LifeGas OXYGEN COMPRESSED UN1072 USP, Distributed by: Linde
Official notice