Luna G3 BPAP System recalled for incorrect therapy pressure data

React Health (3B Medical, Inc.) is correcting Luna G3 BPAP System Model LG3800-30VT devices that may record and report incorrect therapy pressure data when the Targeted VT feature is enabled. Affected devices may not save the actual pressure delivered to the patient to the SD card or send it to React Health Connect. Clinicians who rely on that inaccurate history could adjust settings in ways that delay therapy.

From the FDA notice · Read the official notice ↗

What to do

  1. Contact your health care provider or DME promptly to assess and update your prescribed therapy settings as needed.
  2. Do not change your prescribed settings or stop respiratory therapy without guidance from your health care provider.
  3. Continue using the Luna G3 BPAP System as prescribed until your health care provider gives you additional guidance.
  4. Complete the Customer Recall Notification Response Form from the October 8, 2026 notice and email it to 3B-Recall@reacthealth.com within seven days to get the firmware correction. If you got the device through a DME, contact them to install it.
  5. Clinicians and DME providers: immediately discontinue Targeted VT on affected devices where it is enabled, and coordinate any needed therapy transition.

Is yours affected?

Compare your label or packaging with the recalled product.

Sold in
Nationwide
Made for
Adults (0 months and older)

Details

Illnesses reported
Yes

In the FDA's words

Voluntary Recall – Targeted Pressure Data Reporting Correction – Originally Communicated October 8, 2026

Update Published: November 17, 2026

React Health (3B Medical, Inc.) is providing notice of a voluntary medical device correction (recall) involving specific Luna G3 BPAP System (Model LG3800-30VT) devices due to a data handling issue that may result in incorrect therapy pressure statistical data when the device is operated with Targeted VT enabled.

Read the full announcement

This notice provides important information for durable medical equipment (DME) providers, hospitals, healthcare facilities, clinicians, and patients using affected devices.

Affected Product Identification and Distribution

Product name Luna® G3 BPAP System

Model LG3800-30VT

UDI-DI(s) 06948538365731

Affected serial number(s) APPENDIX A

Number of affected devices 7,533 devices distributed

Distribution locations / recipients US and Puerto Rico

Explanation of the Issue

React Health is recalling specific units of the Luna G3 BPAP System (Model LG3800-30VT) due to a data storage and transmission error while Targeted VT is enabled. Devices operating with Targeted VT enabled will not record the actual pressure statistical data delivered to the patient on the internal SD card or transmit the pressure to the cloud-based management system (React Health Connect).

Users that have not enabled the Targeted VT feature on their Model LG3800-30VT device will not experience this issue.

Potential Risk and Clinical Considerations

The hazardous condition requires the device to be operated with Targeted VT enabled and monitored remotely via either data downloaded to the SD card or via wireless communication to the cloud-based management system (React Health Connect). The consequence of the inaccurate pressure data in historical therapy reports could lead a clinician to adjust device settings based on inaccurate therapy pressure data, which could result in delayed therapy until the setting is corrected. Design features that reduce the risk of this issue include the device adjusting pressure delivered to the patient within the clinician defined minimum and maximum pressures delivering a tidal volume equal to the clinician set tidal volume, within tolerance.

To date, React Health has received fifteen (15) complaints with two (2) serious adverse reports associated with this Luna G3 BPAP System (Model LG3800-30VT) Targeted VT enabled issue.

Clinicians should be aware that inaccurate historical pressure data may lead to misinterpretation of therapy performance, adjustments to device settings based on inaccurate information, or delays in recognizing opportunities to optimize patient therapy. Clinicians should consider these limitations when reviewing historical therapy reports and determining whether additional patient assessment or therapy adjustments are appropriate.

Actions for Patients/Users

Patients using a Luna G3 BPAP System (Model LG3800-30VT) Targeted VT enabled feature should do the following:

Immediately contact your health care provider/physician and/or DME to assess and update prescribed therapy settings as needed .

Do not independently change your prescribed therapy settings or discontinue respiratory therapy without guidance from your health care provider . Your healthcare provider can determine whether your device is affected, evaluate your therapy needs, and coordinate any necessary corrective action or alternative therapy support.

Continue using the Luna G3 BPAP System as prescribed until you receive additional guidance from your health care provider.

Complete the ‘Customer Recall Notification Response Form’ included in the React Health Recall Notification dated October 8, 2026, and send it to React Health within seven (7) days via e-mail to: 3B-Recall@reacthealth.com , for installing firmware correction.

If you have obtained your Luna G3 BPAP System (Model LG3800-30VT) through a DME provider, contact your DME to install the firmware correction.

Actions for Healthcare Professionals and DME Providers

React Health requests that DME providers, hospitals, healthcare facilities, and physicians/clinicians review this information and follow the corrective actions outlined in the original customer recall notification dated October 8, 2026, including as below:

Discontinue Targeted VT : Immediately discontinue use of Targeted VT if enabled on affected Luna G3 BPAP System (Model LG3800-30VT) devices. Healthcare providers should coordinate any necessary therapy transition to maintain appropriate patient support.

Inspect affected inventory : Inspect current s