Ricon wheelchair lifts recalled for possible unintended movement
Ricon S-Series and K-Series wheelchair platform lifts may have a faulty electrical contactor that prevents the controls from stopping the lift. This means a lift could continue moving while someone is on it, creating an injury risk.
From the NHTSA notice · Read the official notice ↗
What to do
- Contact Ricon customer service at 1-800-322-2884 if you own one of these lifts.
- Dealers will install two diodes to fix the wiring at no charge.
- Check your mail for an owner notification letter sent June 29, 2026.
Is yours affected?
Compare your label or packaging with the recalled product.
- Product
- Wheelchair Lift May Inadvertently Move
Details
- Units affected
- 1220
- Recall number
- 26E038000
In the NHTSA's words
- Description
- Ricon Corporation (Ricon) is recalling certain S-Series and K-Series wheelchair platform lifts. The electrical contactor may fail, preventing the controls from stopping the lift. A wheelchair lift that continues to move while occupied increases the risk of injury. Dealers will correct the wiring by installing two diodes, free of charge. Owner notification letters were mailed June 29, 2026. Owners may contact Ricon customer service at 1-800-322-2884.
- Hazard
- A wheelchair lift that continues to move while occupied increases the risk of injury.
- Remedy
- Dealers will correct the wiring by installing two diodes, free of charge. Owner notification letters were mailed June 29, 2026. Owners may contact Ricon customer service at 1-800-322-2884.