Abbott GmbH
2 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Abbott GmbH and translate it into plain English — free weekly digest, optional same-day alerts.
- 2024-04-10FDA-DeviceClass IIAlinity i Anti-HCV Reagent Kit, List Number 08P0521
Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).
- 2024-04-10FDA-DeviceClass IIARCHITECT Anti-HCV Reagent Kit, List Numbers: a) 1L79-25, b) 1L79-35
Customer complaints were received regarding falsely elevated results for some patient samples. Internal studies with customer return samples confirmed potential interactions that may lead to falsely elevated results when processed as below on the same instrument: " ARCHITECT Syphilis TP (LN 8D06) and/or ARCHITECT AdviseDx SARS-CoV-2-IgG II (LN 6S60) precedes ARCHITECT Anti-HCV testing (LN 1L79). " Alinity i Syphilis TP (LN 07P60) precedes Alinity i Anti-HCV testing (LN 08P05).