- 2025-03-19FDA-DeviceClass IINumerous models of nonsterile hemostatic forceps: (1) REF BH951R, Negus Tonsil Forceps HVY-CVD 190MM; (2) REF BH952R, Negus Tonsil Forceps X-HVY-CVD190MM; (3) REF BH957R, Schnidt Tonsil Forceps SLT-CVD185MM; (4) REF BH959R, Schnidt Tonsil Forceps CVD 185MM; (5) REF BH961R, Schnidt Tonsil FCPSOPEN-RINGCVD 185MM; (6) REF BH963R, Schnidt Tonsil FCPSOPEN-RINGCVD185MM; and (7) REF BH965R, Wilson Tonsil Forceps 190MM.
The forceps have been used in ways not covered by the design resulting in breakage of the clamps.
- 2025-03-19FDA-DeviceClass IINumerous models of nonsterile hemostatic forceps: (1) REF FB458R, Glover ATR.CLAMP CVD 215MM; (2) REF FB459R, Glover ATR.CLAMP CVD 245MM; (3) REF FB461R, Glover ATR.CLAMP 195MM; (4) REF FB462R, Glover ATR.CLAMP 210MM; (5) REF FB469R, Leland-Jones Peripheral CLMPSTR 195MM; (6) REF FB567R, Cooley Pediatric Clamp 30DG30/165MM; (7) REF FB568R, Cooley Pediatric Clamp 60DG 30/160MM; (8) REF FB569R, Cooley Pediatric Clamp 90DG30/150MM; (9) REF FB729R, Cooley Pediatric Clamp 17MM 170MM; (10
The forceps have been used in ways not covered by the design resulting in breakage of the clamps.
- 2025-03-19FDA-DeviceClass IINumerous models of nonsterile hemostatic forceps: (1) REF BH198R, Fuchsig Artery Forceps CVD.190MM; (2) REF BH304R, Pean Artery Forceps Straight 140MM; (3) REF BH813R, Artery Forceps Curved 250MM; (4) REF BJ016R, Diss.A.Ligat.Forceps S-Shaped CVD.210MM; (5) REF BJ020R, Overholt-Geiss Suture FCPS #0CVD195MM; (6) REF BJ021R, Overholt-Geiss Suture FCPS #1CVD205MM; (7) REF BJ022R, Overholt-Geiss Suture FCPS #2CVD220MM; (8) REF BJ023R, Overholt-Geiss Suture FCPS #3 CVD225MM; (9) REF BJ025R
The forceps have been used in ways not covered by the design resulting in breakage of the clamps.
- 2025-03-19FDA-DeviceClass IINumerous models of nonsterile hemostatic forceps: (1) REF BH100R, Carrel Artery Forceps 85MM; (2) REF BH104R, Hartmann Mosquito FCSDEL STR 100MM; (3) REF BH105R, Hartmann Mosquito FCPSDELCVD100MM; (4) REF BH108R, Micro-Halsted Forceps DEL STR125MM; (5) REF BH109R, Micro-Halsted Forceps DEL CVM 125mm; (6) REF BH114R, Hartmann Mosquito FCPSDELSTR1X2100MM; (7) REF BH115R, Hartmann Mosquito FCPSDELCVD1X2100MM; (8) REF BH118R, Micro-Halsted FCPS Del STR 1X2125MM; (9) REF BH119R, Micro-Halste
The forceps have been used in ways not covered by the design resulting in breakage of the clamps.
- 2025-03-19FDA-DeviceClass IINumerous models of nonsterile hemostatic forceps: (1) REF BJ500R, Phaneuf Hysterectomy Forceps STR 210MM; (2) REF BJ501R, Phaneuf Hysterectomy Forceps ANG 205MM; (3) REF BJ511R, Holzbach Hyst Forceps CVD245MM; (4) REF BJ520R, Heaney Hyst Forceps Grooved 205MM; (5) REF BJ521R, Heaney Hyst Forceps Grooved 250MM; (6) REF BJ522R, Heaney Hyst Forceps Grooved 195MM; (7) REF BJ524R, Heaney-Rezek Fcpsstr Ser Lgth 210MM; (8) REF BJ527R, Wiener Hyst Forceps Grooves 240MM; (9) REF BJ530R, Heaney
The forceps have been used in ways not covered by the design resulting in breakage of the clamps.
- 2024-06-19FDA-DeviceClass IIAeos Robotic Digital Microscope, Product Code: PV010
Possibility of the robotic arm to drop more than 10cm from its original position after pressing the emergency stop button due to failure of the integrated safety mechanism.
- 2024-03-13FDA-DeviceClass IIDISP.HASSON TROCAR 12/110MM, Product Code EK240SU. For use in laparoscopic procedures.
The sterile blister packaging may be damaged, and sterility may be compromised.
- 2024-03-13FDA-DeviceClass IIDISP.TROCAR W.DILATING PIN 12/110MM, Product Code EK236SU. For use in laparoscopic procedures.
The sterile blister packaging may be damaged, and sterility may be compromised.
- 2024-03-13FDA-DeviceClass IIDISP.TROCAR W.DILATING PIN 12/110MM, Product Code EK234SU. For use in laparoscopic procedures.
The sterile blister packaging may be damaged, and sterility may be compromised.
- 2024-03-13FDA-DeviceClass IIDISP.HASSON TROCAR 10/110MM, Product Code EK230SU. For use in laparoscopic procedures.
The sterile blister packaging may be damaged, and sterility may be compromised.
- 2024-03-13FDA-DeviceClass IIDISP.TROCAR W. DILATING PIN 10/110MM, Product Code EK224SU. For use in laparoscopic procedures.
The sterile blister packaging may be damaged, and sterility may be compromised.
- 2024-02-07FDA-DeviceClass IIMB215R - MILLS DUROGRIP TC Micro Needle Holder, straight, 8:, (203 mm), "TC MILLS NDLE HOLDRRND HNDLSTR203MM"
Needle Holder MB215R was inadvertently distributed as MB362R. Similarly, MB362R was distributed as MB215R during the affected distribution date range. The issue may lead to a short delay in procedure.
- 2024-02-07FDA-DeviceClass IIMB362R - JACOBSON DUROGRIP TC Micro Needle Holder, straight, 8 3/4", (220 mm), "TC JACOBSON NDL HLD STR JAW 220MM";
Needle Holder MB215R was inadvertently distributed as MB362R. Similarly, MB362R was distributed as MB215R during the affected distribution date range. The issue may lead to a short delay in procedure.