- 2024-09-04FDA-DeviceClass IIDR 800. Digital Radiography X-ray System.
Potential for the front lever chain of the DR 800 table to fail.
- 2024-04-10FDA-DeviceClass IIAgfa DX-D 100 system-Digital Radiography mobile X-ray System. A mobile X-Ray modality, consisting of X-Ray console, generator, tube, digital detector, and workstation. This modality is used to generate projection X-Ray images of human patients. Models: 5411/ 050, A5411/ 0300 and A5411/ 0400
Steel cable installed inside the mobile column which supports the weight of the telescopic arm with the tube head has potential risk -an interference could cause the safety system which blocks the arm if the cable is cut off does not work properly, may result in patient injury