Canon Medical System, USA, INC.
8 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Canon Medical System, USA, INC. and translate it into plain English — free weekly digest, optional same-day alerts.
- 2026-02-04FDA-DeviceClass IIAlphenix INFX-8000V, interventional fluoroscopic x-ray system
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
- 2026-02-04FDA-DeviceClass IIAlphenix INFX-8000H, interventional fluoroscopic x-ray system
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
- 2026-02-04FDA-DeviceClass IIAlphenix INFX-8000F, interventional fluoroscopic x-ray system
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
- 2026-02-04FDA-DeviceClass IIAlphenix INFX-8000C, interventional fluoroscopic x-ray system
It has been discovered that there is a possibility that the fixing screws of the ceiling movement gear in the ceiling suspended arm that are part of the X-ray interventional systems may become loose. As a result, ceiling lateral movement may no longer be possible, abnormal noise may be generated from the suspension system and a sensor error may be displayed
- 2026-01-21FDA-DeviceClass IIVANTAGE TITAN 3T Model MRT-3010/MEXL-3010
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
- 2026-01-21FDA-DeviceClass IIVANTAGE GALAN 3T Model MRT-3020/MEXL-3020
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
- 2025-11-05FDA-DeviceClass IIVANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.
Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.
- 2025-10-15FDA-DeviceClass IIThis device is a digital radiography/fluoroscopy system used in a diagnostic and interventional angiography configuration. The system is indicated for use in diagnostic and angiographic procedures for blood vessels in the heart, brain, abdomen and lower extremities.
It has been found that when SPOT Fluoro is first turned on, the default ROI is selected and the collimator blades move to the boundary of the ROI, and the static image is overlayed around the ROI, covering the collimated area. If the customer changes the ROI, the collimator blades and static images are adjusted accordingly. However, when the acquisition program is changed, the collimator returns to the default ROI position which may be larger than the specified one, leaving the X-ray area partially covered by the static image. This is a software problem with the SPOT Fluoro program and will not occur if the function is not used.