- 2023-08-16FDA-DeviceClass IIellacor System with Micro-Coring Technology-Indicated for use by medical professionals for the treatment of moderate and severe wrinkles in the mid and lower face in adults aged 22 years or older with Fitzpatrick skin types I-IV. Model #: Gen 4
Potential failure of a bearing adhesive joint that can occur due to an assembly issue of the X/Y stage of the handpiece. If a broken glue bond were to occur, it could cause an overlap of cores, potentially leading to prolonged healing and irregularities in patient skin texture post-treatment, and/or scarring
- 2023-06-07FDA-DeviceClass IICytrellis ellacor System with Micro-Coring Technology-Dermal Micro-Coring System System intended for the treatment of moderate and severe wrinkles in the mid and lower face. Model #: Gen 4
Irregular core patterns could cause potential overlap of cores at the 8% setting, potentially leading to prolonged healing, irregularities in patient skin texture post-treatment, and/or scarring.