- 2024-03-06FDA-DeviceClass IISchiff Reagent, Hotchkiss-McManus, packaged in an amber glass bottle with a black polypropylene cap.
This IVD is colorless. A customer complaint was received noting it appeared pink. A review of inventory confirmed additional stock appeared pink.
- 2023-12-06FDA-DeviceClass IILugol s Iodine Solution 500ML, Part Number 624-71
Due to the incorrect packaging utilized with lot 3124, there' s an improper seal between the cap and bottle. With an improper seal the Assay (current specification of 24.8 to 27.2 mL) is expected to continue to drop over time as iodine vapors continue to dissipate from the bottles.