15 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Fresenius Kabi USA, LLC and translate it into plain English — free weekly digest, optional same-day alerts.
- 2025-08-06FDA-DeviceClass IILVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Dual Y-Site (Qty 25). Product Code: SET-0032-25.
Potential for external cassette leaks
- 2025-08-06FDA-DeviceClass IILVP Primary Administration Set, Single Outlet Low Sorbing (Qty 25). Product Code: SET-0021-25.
Potential for external cassette leaks
- 2025-08-06FDA-DeviceClass IILVP Blood Products Administration Set Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter (Qty 20). Product Code: SET-0014-20.
Potential for external cassette leaks
- 2025-08-06FDA-DeviceClass IILVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Y-Site (Qty 25). Product Code: SET-0013-25.
Potential for external cassette leaks
- 2025-06-18FDA-DeviceClass IIVENIX INFUSION SYSTEM LVP Blood Products Administration Set, Dual-Inlet, Low-Sorbing, Y-Site, Mesh Filter, 15 mL. Intravascular Administration Set.
Mis-assembly error of Blood Products Administrations Sets where the 200-micron drip chamber filter may be incorrectly positioned and unable to filter out large blood particulates.
- 2025-02-12FDA-DeviceClass IIvenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.
Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.
- 2025-01-15FDA-DeviceClass IIvenix Infusion System (IIS), Large Volume Pump, Model Number: LVP-0004
Potential pneumatic valve failure that may cause the LVP to stop an active infusion and/or could prevent the pump from being used.
- 2024-10-02FDA-DeviceClass IIvenix Infusion System (IIS), LVP Software; LVP-SW-0004, Software Version 5.9.1 and prior. Infusion pump software for infusion management system.
The software has anomalies that have the potential to cause alarms, nonfunctioning pump, or unresponsive screen while continuing therapy. These could cause serious patient harm or death.
- 2024-09-11FDA-DeviceClass ILVP Primary Administration Set, Dual-Inlet, Low-Sorbing, Needle-Free Port, Y-Site (Qty 25), Product Code: SET-0013-25. for intravenous infusion.
A defect may cause an uncontrolled medication flow condition which may result in a dose greater than prescribed for the patient (i.e., medication overdose).
- 2024-04-17FDA-DeviceClass ILVP Software of the Ivenix Infusion System (IIS), Version 5.8.0.
Software has anomalies that have the potential to cause underdose, overdose, or delay in therapy which could lead to serious patient harm or death.
- 2024-03-20FDA-DeviceClass IILVP software of the Ivenix Infusion System (IIS); LVP-SW-0004
Retroactively reported corrections from 2023: 1) A software defect may cause an incorrect (Fail-Stop) alarm when an administration set is loaded or coupled while the pump is executing the power-up sequence. May lead to delay in therapy. 2) Alert is not annunciated informing the clinician that the bolus cannot be delivered when the entered bolus dose exceeds the Care Profile Hard Rate Max limit and Rapid Bolus is selected. May lead to over infusion. Both issues were resolved in all fielded product in software version 5.8.0, which was installed in affected units May thru August 2023.
- 2024-03-20FDA-DeviceClass IILVP software of the Ivenix Infusion System (IIS), LVP-SW-0004
Retroactively reported correction from 2022: A software defect may cause an incorrect (Fail-Stop) Alarm when an administration set is loaded with an upstream occlusion present, resulting in a non-recoverable (Fail-Stop) Pump Problem alarm. This may lead to delay of therapy, and patient may experience from no to minimal to serious harm. This issue was resolved in all fielded product in software version 5.2.1, which was installed in September 2022.
- 2023-12-27FDA-DeviceClass IIvenix Infusion System (IIS), Large Volume Pump LVP-0004
The device may experience mechanical interference on the Fluid Valve pins, which will trigger a Pump Problem alarm. The issue may lead to delay or interruption of therapy. Depending on the therapy and duration of delay/interruption, the issue may lead to serious harm or death.
- 2023-10-04FDA-DeviceClass IIIvenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.
Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set
- 2023-04-19FDA-DeviceClass IIvenix Infusion System (IIS), Large Volume Pump, Model No. LVP-0004
Fluid ingress that can cause a loss of electrical funtion and failure of the Set ID Sensory, resulting in the prevention of infusion or halt of infusion.