Fresenius Medical Care Holdings, Inc.
41 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Fresenius Medical Care Holdings, Inc. and translate it into plain English — free weekly digest, optional same-day alerts.
- 2025-10-15FDA-DeviceClass IIOptiflux¿ High Flux E-beam Dialyzer Model Numbers: (1) 0500316E OPTIFLUX F160NRe 12/CS 1.5SA EBEAM, (2) 0500318E OPTIFLUX F180NRe 12/CS 1.7SA EBEAM, (3) 0500320E OPTIFLUX F200NRe 12/CS 1.9SA EBEAM, (4) 0500325E OPTIFLUX F250NRe 12/CS 2.5SA EBEAM
To provide further information to the user to ensure proper use of the medical device. the caps had been changed from a threaded version to a press-on version whereby the user must press the caps firmly onto the dialyzer to securely affix the cap.
- 2025-07-02FDA-DrugClass II0.9% Sodium Chloride Injection, USP, Flexible Plastic Container, Rx Only Fresenius Medical Care North America, Waltham, MA 02451, 1-800-323-5188, NDC 49230-300-10
Lack of Assurance of Sterility
- 2025-06-25FDA-DeviceClass II5008X HDF Hemodialysis Blood Tubing Set, 24 units per case, with the following models: a. 5008X Standard HDF Bloodline, Model Number: 03-5100-7. b. 5008X Standard HDF Bloodline w/CLIC, Model Number: 03-5100-7C. c. 5008X Standard HDF Bloodline w/ Twister, Model Number: 03-5150-2.
Potential of blood leak occurring between the lower port of the venous chamber and the tubing include: hemodiafiltration (HDF), hemodialysis (HD), hemofiltration (HF), and isolated ultrafiltration.
- 2025-01-15FDA-DeviceClass IIDialyzer Optiflux 160NRe
Potential for internal blood leaks due to cracked polyurethane
- 2024-09-18FDA-DeviceClass IIBlood Pressure Module BPM3 TM-2917B (Spare Part Numbers 191378 or 191389) with firmware version S005 installed in 2008 Series Hemodialysis Machines with Part Numbers 190573, 190713, 190766, 190858, 190895, 190908, 191124, 191126, 191128, and 191130
Machines equipped with BPM3 Blood Pressure Modules (Blood pressure monitor TM-2917B with firmware version S005) will not record Blood Pressure measurements and does not alarm if the cuff tubing is kinked close to the cuff.
- 2024-07-24FDA-DrugClass IIDELFLEX PERITONEAL DIALYSIS SOLUTION With LOW MAGNESIUM / LOW CALCIUM 4.25% DEXTROSE and attached stay "safe Exchange Set, 2500mL (Approx. 50 mL excess), Single Dose Container Sterile and Non-Pyrogenic, For Intraperitoneal Administration Only, Fresenius Medical Care NA Waltham, MA 02451, 1-800-323-5188 NDC 49230-212-94
This product is being recalled due to the tube weld failure presents itself as a slow leak and can be difficult to detect.
- 2024-07-17FDA-DeviceClass IIRTLR190395: 2008K@HOME MACHINE, SHORT CAB,OLC/DP,HP - Blood Pump Rotor spare part, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- 2024-07-17FDA-DeviceClass IIRTL190904: 2008K@HOME HEMODIALYSIS SYSTEM W/ BIBAG - Blood Pump Rotor spare part, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- 2024-07-17FDA-DeviceClass IIRTL190828: 2008K@HOME HEMODIALYSIS SYSTEM - Blood Pump Rotor spare part, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- 2024-07-17FDA-DeviceClass IIRTL190395: 2008K@HOME MACHINE, SHORT CAB,OLC/DP,HP - Blood Pump Rotor spare part, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- 2024-07-17FDA-DeviceClass II190395: 2008K@HOME MACHINE, SHORT CAB,OLC/DP,HP - Blood Pump Rotor spare part, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- 2024-07-17FDA-DeviceClass II191014: 2008K@HOME HD SYSTEM W/BIBAG (CANADA) - Blood Pump Rotor spare part, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- 2024-07-17FDA-DeviceClass II190828: 2008K@HOME HEMODIALYSIS SYSTEM - Blood Pump Rotor spare part, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- 2024-07-17FDA-DeviceClass II190904 2008K@HOME HEMODIALYSIS SYSTEM W/ BIBAG - Blood Pump Rotor spare part, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- 2024-06-19FDA-DeviceClass IIUS AquaA 900H-Typ USA-3 x 208v-Intended for the purification of water to be used for hemodialysis Vivonic Part Numbers: G02020110 Fresenius Part Number: 24-090A-1
Concentrate Flow was exceeding the maximum 19.9 bar set point. The standard default set point range is between 4 and 18.Error message (F020104, "concentrate pressure alarm limit exceeded") concentrate pressure is too high and stops the dialysis water supply.
- 2024-05-29FDA-DeviceClass II2008T BlueStar Hemodialysis Machine, Model 191130 with Component: Blood Pump Rotor, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- 2024-05-29FDA-DeviceClass II2008T BlueStar Hemodialysis Machine, Model 191128 with Component: Blood Pump Rotor, Model Number: F40015481 Rev A.
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- 2024-05-29FDA-DeviceClass II2008T BlueStar Hemodialysis Machine, Model 191126 with Component: Blood Pump Rotor, Model Number: F40015481 Rev A.
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- 2024-05-29FDA-DeviceClass II2008T BlueStar Hemodialysis Machine, Model 191124 with Component: Blood Pump Rotor, Model Number: F40015481 Rev A.
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- 2024-05-29FDA-DeviceClass IIBlood Pump Rotor, Model Number: F40015481 Rev A
An increased number of complaints were reported on the Blood Pump Rotor (P/N - F40015481 REV A) after a period of use by clinics noting loose/dislodged guide sheaves.
- 2024-05-22FDA-DeviceClass IINovalung sensor box is a component of the Novalung Console (F30000162)-Indicated for long-term (> 6 hours) respiratory/ cardiopulmonary support that provides assisted extracorporeal circulation and physiologic gas exchange (oxygenation and CO2 removal) of the patient's blood in adults with acute respiratory failure or acute cardiopulmonary failure Product Code: F30000163
Issue related to Novalung sensor box, which is a component of the Novalung Console (Part Number F30000163). In certain instances, error messages #206 (yellow) and #208 (red) technical failure, flow measurement during use of the Novalung system.
- 2024-04-03FDA-DeviceClass Istay-safe /Luer Catheter Ext. 6 in
The peroxide cross-linked tubing has leachables identified as NDL PCBAs.
- 2024-04-03FDA-DeviceClass Istay-safe /Luer Lock Catheter Ext. 18 in
The peroxide cross-linked tubing has leachables identified as NDL PCBAs.
- 2024-04-03FDA-DeviceClass Istay-safe /Luer Lock Catheter Ext. 12 in
The peroxide cross-linked tubing has leachables identified as NDL PCBAs.
- 2024-04-03FDA-DeviceClass Istay-safe /Luer Lock Adapter 4 in
The peroxide cross-linked tubing has leachables identified as NDL PCBAs.
- 2024-04-03FDA-DeviceClass Istay-safe /Safe Lock Catheter Ext. 12 in
The peroxide cross-linked tubing has leachables identified as NDL PCBAs.
- 2024-01-10FDA-DrugClass IIDELFLEX Peritoneal Dialysis Solution in Biofine container 1.5% Dextrose, packaged in 6000mL bags, Rx only, Fresenius Medical Care NA Waltham, MA 02451, NDC 49230-206-62.
Lack of Sterility Assurance
- 2024-01-03FDA-DeviceClass IIILiberty Select Cycler, Model numbers RTLR180343 and RTLR180111
Affected Liberty Select cyclers may display an incorrect time stamp on the treatment data report sent to the clinic.
- 2023-12-13FDA-DeviceClass IIStay Safe Cap
Updating the labeling to include additional warnings and contraindications for potential side effects.
- 2023-10-18FDA-DeviceClass I2008T HD SYS. W/O CDX W/BIBAG BLUESTAR
Potential PCBA leaching from tubing of hemodialysis machines.