Landauer
40 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Landauer and translate it into plain English — free weekly digest, optional same-day alerts.
- 2024-03-27FDA-DeviceClass IImicroStar Reader, Model Numbers: 10015-000 (marketed in EMEA as 100, 102, 110, and 201), 10016-000 (marketed in EMEA as 101), 10036-000, 10044-000, 10055-000, BC30058, and BC30059
Permanent discontinuance of the microSTARii medical dosimetry system. Landauer has initiated this action to remove all reference to use of the microStar readers with the nanoDot or any otherbdosimeter for medical applications from the product s user manual and software and to remove microStar readers from customers who use them only for medical applications.
- 2024-01-03FDA-DeviceClass IILandauer microSTARii reader, Medical Dosimetry System, Model Numbers: 18000-000/18001-000/18007-000
MicroSTARii readers may be contributing to the nanodot dosimeter measurement inaccuracy in two possible ways. The first is related to the interaction of the reader s LED beam profile with the nanoDot dosimeter. The second is debris formation due to drawer actuation, resulting in a change of reader response that may not be detectable unless known dose QCs are being run regularly.
- 2023-09-20FDA-DeviceClass IIReader, nanoDot adapter, EU Model Number VINLADA003; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IINanoDots, Model Number 04297-000; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IINanoDots, Model Numbers: a) BC30023, description: QC Set For Microstar Dots; b) BC30083, description: nanoDot QC Set - 80kVp - Sale; c) BC30084, description: nanoDot QC Set - 80kVp - No Charge; d) BC30088, description: nanoDot Calibration Set - 80kVp - No Charge; e) BC30095, description: Dot, D1DNN, in carrier sealed in bag; f) BC30110, description: Calibration set, nanoDot; g) BC30141, description: nanoDot QC Set - Cs137 - Sale; h) BC30142, description: nanoDot Calibrat
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IIConstancy (Cs-137) nanoDot D2DNN, Model Numbers: a) 18155-000 (EMEA number VKITCON001); b) 18155-KIT; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IIConstancy (80 kVp) nanoDot D2DNN, Model Numbers: a) 18150-000 (EMEA number VKITCON002); b) 18150-KIT; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IICalibrate (unexposed) nanoDot D2DNS, Model Numbers: a) 18140-000; b) 18140-KIT; c) 18140-SET; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IIQC (Cs-137) nanoDot D2DNS, Model Numbers: a) 18135-000; b) 18135-KIT; c) 18135-SET; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IIQC (80 kVp) nanoDot D2DNS, Model Numbers: a) 18131-000; b) 18131-KIT; c) 18131-SET; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IIQC (80 kVp) nanoDot D2DNS, Model Numbers: a) 18130-000; b) 18130-KIT; c) 18130-SET; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IICalibrate (Cs-137) nanoDot D2DNS, Model Numbers: a) 18125-000; b) 18125-KIT; c) 18125-SET; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IICalibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 18121-000; b)18121-KIT; c) 18121-SET; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IICalibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 18120-000; b) 18120-KIT; c) 18120-SET; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IInanoDot D2DNN, Model Numbers: a) 18105-000; b) 18105-KIT; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IInanoDot D2DNS, Model Numbers: a) 18100-000; b) 18100-1MO; c) 18100-2MO; d) 18100-KIT; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IInanoDot D2DNN , Model Numbers: a) 04293-000; b) 04293-KIT; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IIQC (Cs-137) nanoDot D2DNS, Model Numbers: a) 04225-000 (EMEA number VINLNAN005); b) 04225-KIT; c) 04225-SET; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IICalibrate (Cs-137) nanoDot D2DNS, Model Numbers: a) 04224-000 (EMEA number VINLNAN006); b) 04224-KIT; c) 04224-SET; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IIQC (80 kVp) nanoDot D2DNS, Model Numbers: a) 04218-000 (EMEA number VINLNAN004); b) 04218-KIT; c) 04218-SET; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IICalibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 04217-000 (EMEA number VINLNAN003); b) 04217-KIT; c) 04217-SET; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IInanoDot C - item 3 of 3 for 03500-000, Model Number 03503-000; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IInanoDot B - item 2 of 3 for 03500-000, Model Number 03502-000; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IInanoDot A - item 1 of 3 for 03500-000, Model Number 03501-000; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IInanoDot kit (contains nanoDot A, B, C), Model Number 03500-000; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IIQC (Cs-137) screened nanoDot D2DNS, Model Number 03067-OTO; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IIQC (80 kVp) screened nanoDot D2DNS, Model Number 03066-OTO; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IICalibrate (Cs-137) screened nanoDot D2DNS, Model Number 03065-OTO; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IICalibrate (80 kVp) screened nanoDot D2DNS, Model Number 03064-OTO; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.
- 2023-09-20FDA-DeviceClass IInanoDot D2DXS, Model Numbers: a) 03063-1MO; b) 03063-3MO; c) 03063-OTO; d) 03063-SMO; radiation monitoring dosimeter used with the microSTAR readers
LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.