Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
120 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) and translate it into plain English — free weekly digest, optional same-day alerts.
- 2023-06-28FDA-DeviceClass IICD-VR VISIA AF S US/OUS IS1/DF1, Model Number DVAC3D1; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass IICD-VR VISIA AF XT OUS DF4, Model Number DDVAB2D4; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass IICD-VR VISIA AF XT OUS IS1/DF1, Model Number DVAB2D1; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass IICD-VR DVAB1D4 VISIA AF US DF4, Model Number DVAB1D4; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass IICD-VR DVAB1D1 VISIA AF US IS1/DF1, Model Number DVAB1D1; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD CROME HF QUAD MRI IS4 DF4, Model Number DTPC2QQ; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD CROME HF QUAD MRI IS4 DF1, Model Number DTPC2Q1; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD CROME HF MRI IS1 DF4, Model Number DTPC2D4; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD CROME HF MRI IS1 DF1, Model Number DTPC2D1; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD COBALT HF QUAD MRI IS4 DF4, Model Number DTPB2QQ ; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD COBALT HF QUAD MRI IS4 DF1, Model Number DTPB2Q1 ; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD COBALT HF MRI IS1 DF4, Model Number DTPB2D4; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD COBALT HF MRI IS1 DF1, Model Number DTPB2D1; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass IICD COBALT XT HF QUAD OUS, Model Number DTPA2QQG; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD COBALT XT HF QUAD MRI IS4 DF4, Model Number DTPA2QQ; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD COBALT XT HF QUAD MRI IS4 DF1, Model Number DTPA2Q1; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD COBALT XT HF MRI IS1 DF4, Model Number DTPA2D4; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD DTPA2D1PX COBALT XT HF QUAD OUS, Model Number DTPA2D1PX; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD DTPA2D1G COBALT XT HF QUAD OUS, Model Number DTPA2D1G; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD COBALT XT HF MRI IS1 DF1, Model Number DTPA2D1; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD DTMC2QQ COMPIA MRI QUAD OUS DF4, Model Number DTMC2QQ; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD DTMC2D4 COMPIA MRI OUS DF4, Model Number DTMC2D4; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD DTMC2D1 COMPIA MRI OUS DF1, Model Number DTMC2D1; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD DTMC1QQ COMPIA MRI QUAD US DF4, Model Number DTMC1QQ; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD DTMC1D1 COMPIA MRI US DF1, Model Number DTMC1D1; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD DTMB2QQ AMPLIA MRI QUAD OUS DF4, Model Number DTMB2QQ; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD DTMB2Q1 AMPLIA MRI QUAD OUS DF1, Model Number DTMB2Q1; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD DTMB2D4 AMPLIA MRI OUS DF4, Model Number DTMB2D4; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD DTMB2D1 AMPLIA MRI OUS DF1, Model Number DTMB2D1; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.
- 2023-06-28FDA-DeviceClass ICRTD DTMB1QQ AMPLIA MRI QUAD US DF4, Model Number DTMB1QQ; Implantable Cardioverter Defibrillators
There is a rare potential for reduced- or no-energy output during high voltage (HV) therapy (typically 0-12J) in implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds) manufactured with a specific (glassed) feedthrough, including currently available ICDs and CRT-Ds.