NOXBOX LTD
3 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for NOXBOX LTD and translate it into plain English — free weekly digest, optional same-day alerts.
- 2025-10-15FDA-DeviceClass INOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I
The unexpected internal fault / system diagnostic error to result when a rapid succession of button selections is made to the user interface without waiting for the device to respond to user prompts.
- 2025-10-15FDA-DeviceClass INOxBOXi Nitric Oxide Delivery System; Model Numbers: NOXBOX-I
Fluctuations may be more likely to occur if the total flow through the device sensor (used by the device to determine flow) is less than the minimum device requirement of 0.5 LPM. In addition, dose fluctuations have been observed when Bunnel LifePulse HFJV system or conventional ventilators have the following device settings: low service pressures generally less than 3.0 PSI, a pressure difference greater than 5.0cm H2O; or at breath per minute rates of 300 (5 Hz).
- 2025-09-17FDA-DeviceClass IINOxBOXi Nitric Oxide Delivery System
Replacement of the internal Sample Pump due to high failure rates.
Get notified about new NOXBOX LTD recalls
Free weekly digest. We email when a new federal recall affects NOXBOX LTD or anything else in your household.
Get the Sunday Brief