Qualigen Inc
2 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Qualigen Inc and translate it into plain English — free weekly digest, optional same-day alerts.
- 2025-01-15FDA-DeviceClass IIIThe FastPack IP Sex Hormone Binding Globulin Immunoassay, Part #25000081, is for the quantitative measurement of SHBG in human serum and plasma. For in vitro diagnostic Use and contains 30 FastPacks.
Three lots of the FastPacks kitted into the FastPack SHBG Complete Kits were labeled with the incorrect expiration date. The following is a listing of affected lots: Lot Number: 2408018-1, 2409015-1, and 2409015-2. The likelihood of patient harm is expected to be low as routine Quality Control should detect kits that are impacted by signal deterioration due to age.
- 2024-10-16FDA-DeviceClass IIFastPack TSH Calibrator Kit, REF: 25000024, For use in conjunction with the FastPack IP TSH Immunoassay and FastPack IP System Analyzer
Thyroid Stimulating Hormone (TSH) calibrator card has an incorrect barcode linked to incorrect information, that may allow successful system calibration, but the TSH results obtained could be higher than expected.
Get notified about new Qualigen Inc recalls
Free weekly digest. We email when a new federal recall affects Qualigen Inc or anything else in your household.