Radiometer Medical ApS
3 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Radiometer Medical ApS and translate it into plain English — free weekly digest, optional same-day alerts.
- 2025-04-30FDA-DeviceClass IIRadiometer ABL90 Series - Radiometer ABL90 FLEX Model Number 393-090 Radiometer ABL90 Series - Radiometer ABL90 FLEX PLUS Model Number 393-092 The Radiometer ABL90 Series is a portable, automated analyzer for use in the laboratory environment, near patients or point-of-care settings.
ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (watchdog) that detects a freeze and automatically restarts the software to continue operating the analyzer. However, even if a restart occurs, any sample being processed at that time will be lost, potential to lead to delay in test results.
- 2024-07-31FDA-DeviceClass IIABL800 FLEX Model Numbers 393-800 and 393-801 The ABL800 FLEX analyzers are intended for: " In Vitro testing of samples of whole blood for the parameters pH, pO2, pCO2, cK+, cNa+, cCa2+, cCl , cGlu, cLac, cCrea, ctBil, and co-oximetry parameters (ctHb, sO2, and the hemoglobin fractions FO2Hb, FCOHb, FMetHb, FHHb and FHbF) " In vitro testing of samples of expired air for the parameters pO2 and pCO2 " In vitro testing of pleura samples for the pH parameter.
Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles. This may result in a probability of reporting biased out-of-specification pH results on blood samples.
- 2023-04-26FDA-DeviceClass IIAQURE REF 933-599 Software Versions 2.5.2 2.5.3 2.5.4 2.6.0 2.6.1 The AQURE system is intended to let you manage analytical devices and operator profiles. The user can associate patient data with test data. The system shows test results. The system receives data from connected devices at the point-of care or laboratory. It can send test results to the HIS/LIS. The system lets you send commands to selected devices. The system uses data related to the performance of devices, to tell user
Due to potential software issue that may result in patient mix-up information.