Roche Diagnostics
2 federal recalls on file. We monitor every CPSC, FDA, USDA, and NHTSA notice for Roche Diagnostics and translate it into plain English — free weekly digest, optional same-day alerts.
- 2026-03-11FDA-DeviceClass IIContact Roche about cobas pro analyzer software issue
Software defect, which allows the system to accept erroneous, non-monotonous calibrations for Spline-type assays. Cobas pro integrated solutions is an automated analyzer, intended for running qualitative, semi-quantitative and quantitative clinical chemistry and immunochemistry assays, as well as ion-selective measurements. When an erroneous calibration is active, the instrument fails to calculate new values and instead repeats the last successfully calculated result from any Spline-type assay for all subsequent measurements, leading to identical and erroneous patient and QC results. The defect could lead to patients receiving erroneous lab results, which could result in a remote risk of serious adverse health consequences. The following assays use spline type calibration and are affected by this issue: Cystatin C Gen.2, Ferritin Gen.4, Lipoprotein (a) Gen. 2, Lipoprotein (a) molarity, Vancomycin Gen.3, Kappa Free Light Chains Partner Channel, Lambda Free Light Chains Partner Channel, and fCAL turbo Partner Channel.
- 2026-01-14FDA-DeviceClass IIContact your lab if you used Roche Elecsys Anti-TSHR
Issues identified: 1) Results vary by Instrument Platform; 2) Calibrator Lot-to-Lot Variability. Issues may lead to delayed or incorrect diagnosis, as well as delay of treatment or incorrect treatment.
Get notified about new Roche Diagnostics recalls
Free weekly digest. We email when a new federal recall affects Roche Diagnostics or anything else in your household.
Get the Sunday Brief